AEJ-C/AES-C/PLJ-C-BA-e-1621479.3.7Manual / automatic data output / AUTO THRESHOLDThis function is used whether and how data output is to take place.Available settings:Menu itemAvailable settings / explanationMODEManual output of stable weighing values afterpressingPRINTkeyManual output for stable and instable weighingvalues after pressing PRINT keySequence of operations:•Set balance to zero•Place goods to be weighed on balance. Thefirst stable weighing value will automaticallybe issued when the current weighing valueexceeds the set threshold value.•A new output requires that the weighingvalue first falls short and then exceeds theset threshold.AUTOTHRESHOLDSetting for threshold in [g], from which automatic data output is tostart.