166 AEJ-C/AES-C/PLJ-C-BA-e-162129.6.19 Enter reference weight (percentage weighing)Command: RM˽XXXXX CR LFPossible replies:RM˽OK CR LF Command executedRM˽I CR LF Command valid but cannot be executed at the moment (e. g.incorrect operating mode)ES CR LF Syntax error (incorrect weight format)29.6.20 Confirm / simulation “Press PRINT-key“Command: SS CR LFCommand SS CR LF is used to save the weighing result automatically to thedatabase and if an optional printer is connected to print out the result.29.6.21 Start internal adjustmentCommand: IC CR LFPossible replies:IC ˽A CR LF Adjustment runningIC ˽D CR LF Adjustment finishedIC ˽E CR LF Command valid but cannot be finished at the moment (e. g .range or time limit for stability display exceeded)IC ˽I CR LF Command valid but cannot be executed at the moment29.6.22 Disable automatic internal adjustmentCommand: IC1 CR LFPossible replies:IC1 ˽OK CR LF Command executedIC1 ˽I CR LF Command valid but cannot be executed at the moment (Notavailable in weighing balances with type approval)IC1 ˽E CR LF Command valid but not available at the moment