AEJ-C/AES-C/PLJ-C-BA-e-1621 16329.6.8 Send weighing value immediately in current weighing unitCommand: SUI CR LFPossible replies:SUI˽I CR LF Command valid but cannot be executed at the moment1 2 3 4 5 6 7-15 16 17-19 20 21S U I Stabilitysign ˽ Signs Weight ˽ Unit CR LFExample:S U I CR LF Sent command, see chap. 29.4S U I ?˽-˽˽˽5 8 . 2 3 7˽k g˽CR LF Command executed. Weighing value is shownin current weighing unit29.6.9 Start continuous output in standard weighing unitCommand: C1 CR LFPossible replies:C1˽I CR LF Command valid but cannot be executed at the momentC1˽A CR LF Command accepted, command executingWeighing values are shown in default weighing unit1 2 3 4 5 6 7-15 16 17-19 20 21S I ˽ Stabilitysign ˽ Signs Weight ˽ Unit CR LF29.6.10 Stop continuous output in standard weighing unitCommand: C1 CR LFPossible replies:C0˽I CR LF Command valid but cannot be executed at the momentC0˽A CR LF Command accepted, command executing