AEJ-C/AES-C/PLJ-C-BA-e-1621 16729.6.23 Enable automatic internal adjustmentCommand: IC0 CR LFPossible replies:IC0 ˽OK CR LF Command executedIC0 ˽I CR LF Command valid but cannot be executed at the moment (Notavailable in weighing balances with type approval)29.6.24 Enter serial numberCommand: NB CR LFPossible replies:NB˽A˽”x“ CR LF Command accepted, stating serial numberNB˽I CR LF Command valid but cannot be executed at the moment”x“: Serial number of device29.6.25 Disable weighing balance keyboardCommand: K1 CR LFPossible replies:K1˽I CR LF Command valid but cannot be executed at the momentK1˽OK CR LF Command accepted, command executing29.6.26 Enable weighing balance keyboardCommand: K0 CR LFPossible replies:K0˽I CR LF Command valid but cannot be executed at the momentK0˽OK CR LF Command accepted, command executing