CHAPTER 1 INTRODUCTION1.4 TEST MASSES REQUIREDThe masses required to test the Discovery Balances must meet or exceed the requirements ofthe Class specified. The calibration points for the Discovery Models are listed in Table 1-1.Use the minimum number of masses to total the calibration point value.Table 1-1. MASSES USED TO TEST DISCOVERY BALANCESSPAN CALIBRATION POINTSWEIGHT CLASS MODEL CAPACITY (g) DEFAULT ALTERNATEASTM OIML CAL POINTS (g) CAL POINTS (g)Class 1 E1 DV114C 110 100 50Class 1 E1 DV214C 210 200 100Class 1 E1 DV314C 310 300 100, 200Class 1 E1 DV215CD 81/210 200 100Discovery Balance Service Manual 1-3 Ohaus Corporation www.ohaus.com