CHAPTER 3 REPAIR PROCEDURES3.2.2 Load Cell Disassembly7. Install the 2 thumb screws from Service Toolset, as shown.Hanger Thumb screwsFigure 3- 40. Thumb screws and Hangeron top of Load Cell.It is now safe to remove the Load Flexure.Load Flexure ScrewsFigure 3-41. Screws securing LoadFlexure to Hanger on top of Load Cell.8. Carefully remove the two screws shown, together with 2 washers each: one steel(closest to screw head) and one aluminum.Discovery Balance Service Manual 3-23 Ohaus Corporation www.ohaus.com