CHAPTER 3 REPAIR PROCEDURES3.2.2 Load Cell Disassembly3. Tip the Load Cell onto the back frame and loosen the 2 screws that connect thebottom Flexure Arm to the frame, NO MORE than half a turn.This is preparation for installing the fixture to protect the Flexure Arms.Screw connectingbottom Flexure Armassembly to frameScrew connectingbottom Flexure Armassembly to frameFigure 3-35. Screws on bottom Flexure Arm assembly.Screw connectingtop Flexure Armassembly to frame Screw connecting topFlexure Arm assemblyto frameFigure 3-36. Screws on top Flexure Arm assembly.4. Loosen the 2 screws that connect the top Flexure Arm assembly to the frame,NO MORE than half a turn. This is preparation for mounting the fixture to protect theFlexure Arms.Discovery Balance Service Manual 3-21 Ohaus Corporation www.ohaus.com