CHAPTER 3 REPAIR PROCEDURES3.2.5 Installing the Load Cell4. Tighten the 3 mounting screws after aligning Load Cell. Tighten screw at point oftriangle first, then re-check alignment, and adjust as necessary. Then tighten 2screws at base of triangle. Re-check alignment and adjust as necessary.5. Connect the flexible cable on Load Cell, using the back of a pair of tweezers to presscable into place. Also connect cable on Calibration Motor Assembly.6. Place the Load Cell Cover over the Load Cell.Figure 3-93. Replacing Load Cell Cover.7. Note the notches for cables: top-left for Load Cell cable, and bottom-right forCalibration Motor Assembly cable.Figure 3-94. Load Cell with Cover in proper position,cables resting in notches (top-left and bottom-right).Ohaus Corporation www.ohaus.com 3-46 Discovery Balance Service Manual