EN-22 ExplorerProADJUSTMENTSThe G/N/T WEIGHING display contains three soft keys at the topof the screen UNITS, 2nd UNITS and TARE.TAREWhen the TARE soft key is selected, pressing the Enter button willstore the current display reading as the tare value.2nd UNITSWhen 2nd UNITS soft key is highlighted, each press of the Enterbutton changes the measuring units that are on and will appearunder the large numerals. The last option when cycling throughthe units is Off.UNITSWhen the UNITS soft key is highlighted, each press of the Enterbutton cycles through all active measuring units on the Gross,Net and Tare displays.3.4.7 FillingFilling permits entering a target reference weight. Material can beplaced on the pan and by monitoring the Difference weight andthe bar display, an accurate fill can be achieved. A split bardisplay indicates 0% to 90% on the first bar and 90% to 110%on the second bar.FI LL ING 01/01/03 03:05:45 PMTA RGET S ET U P NEW TA R GET0.00 gTarget: 50.00 gDiff: -50.00 gPct: 0%0.00 g*0%50% 90%110%100%ADJUSTMENTSThe FILLING display contains three soft keys at the top of thescreen TARGET, SETUP and NEW TARGET.NEW TARGETWith the NEW TARGET soft key highlighted, press the Enter button.Follow display instructions to “Add Target Weight, Press Enter”.This establishes a new target weight and is displayed on thescreen.SETUPWhen SETUP soft Key is highlighted, press the Enter button. TheFILLING SETUP display is shown.FI LLING S ET UPUnit : g2 nd unit : gTarget Weight : 0000Display S ettings : CustomTarget Key : UnlockedNew Target Key : UnlockedExitUnits: g, kg, etc...(default is g)Cycles through all units turned On in the Units menu.2nd UNITS: g, kg, etc...(default is g)Cycles through all units turned On in the Units menu.Target Weight: 0-99999999 (default is 0)Allows entering a specific target filling weight.Display Settings: Custom, Default (default is Default)Setting the display to Custom allows individual items TargetWeight, Difference, Percent, 2nd weight to be turned On or Off. Inthe default setting, all items are turned on.Target Key: Locked, Unlocked (default is Unlocked)This feature allows to lock or unlock the soft key function.New Target Key: Locked, Unlocked (default is Unlocked)This feature allows to lock or unlock the soft key function.Exit:When selected, balance returns to the Filling Mode.USEFILLINGA sample (target weight) is placed on the pan and the Enterbutton is pressed. The balance stores this weight as shown onthe display as Target. The sample is removed from the pan andthe material is added to the pan. The display provides TargetWeight, Difference weight, Percentage and the split bar at thebottom of the screen displays the percentage of the samplecompared against the target weight. When a container is used,zero the balance before placing the actual sample on the pan.See Adjustments to set other balance parameters.