EN-28 ExplorerPro3.5.7 Print OptionsPress the Menu button, and select Print Options.Press the Enter button PRINT OPTIONS screen is displayed.P R INT OPTIONSAuto Print : OffPrint Interval : OffS table Data : OffNumeric Data : OffExitAuto Print: Off, interval, etc... (default is Off)When enabled, the Auto Print feature causes the balance toautomatically output display data in one of three ways:continuously, at user specified time intervals, or upon stability.OFF Turns off the auto print feature.Interval Provides a user specified printing intervalStable Provides printed data each time a stablereading is achieved.Continuous Outputs data continuously.Print Interval: 1-3600 (default is 1)When Auto Print is set to Interval, a specified printing intervalbetween 1 and 3600 seconds can be set.Stable Data: Load, Load & Zero (default is Load)When Auto Print is set to Stable, an option of Load or Load &Zero can be set.Load Will print stable load data only.Load & Zero Will print stable load data and stable zerodata.Numeric Data: On, Off (default is Off)When set On, will print numeric data.3.5.8 GLP Print OptionsGLP stands for Good Laboratory Practice. The GLP Print Optionsenables the printing of a Date & Time, Balance ID, Project Name,User Name, Calibration, Reference, Application Mode and Resultlines. Press the Menu button, and select GLP Print Options.Press the Enter button GLP PRINT OPTIONS screen is displayed.G L P PRINT OPTION SProject Name : OHA U SUser Name : OHA U SDate & Time : OffBalance ID : OffProject Name : OffUser Name : OffCalibration : OffR eference : OffApplication Mode : OffR esult L ine 1 : OffR esult L ine 2 : OffR esult L ine 3 : OffProject Name: (default is OHAUS)A project name of up to 8 characters may be entered.User Name: (default is OHAUS)A user name of up to 8 characters may be entered.Date & Time: On, Off (default is Off)When set On, will output date and time.Balance ID: On, Off (default is Off)When set On, will output Balance ID.Project Name: On, Off (default is Off)When set On, will output project name.User Name: On, Off (default is Off)When set On, will output User name.Calibration: On, Off (default is Off)When set On, prints out the calibration data after the completionof every calibration process. See sample of span calibration.---------------SPAN CAL--------------Cal: 400.000 gOld:399.445 gDiff:-0.555 gWt. ref______________________Name_______________________-----------------END--------------------Reference: On, Off (default is Off)When the Reference function is set ON, it will output the value ofweight used as a reference in either Percent Weighing mode orParts Counting mode.Application Mode: On, Off (Default is Off)When set to On, the Application name will be printed during aprint operation.SAMPLE PRINTOUT