EN-34 ExplorerProCommandCharacter DescriptionC Begin span calibrationxD Set 1 second print delay (set x = 0 for OFF, or x = 1 for ON)PM Application mode.xFL Set Averaging Filter Level 1= Low, 2= Medium, 3= HighL Begin linearity calibrationP Print displayed weight (stable or unstable)T Same as pressing Tare key.PV Version: print name, software revision and LFT ON (if LFT is set ON)xAL Set Auto-Zero level to x. x = 0 for OFF, x = 1 for 0.5d, x=2 for 1.0d, x=3 for 2.0d, x=4 for5.0d.Esc R Resets Setup and Print menus to factory defaultsOn Turns balance onOff Turns balance offx% Set % reference weight (x) in current unitx# Set PC reference weight (x) in current unitP# Print PC reference weightP% Print percent reference weightxM Set current Application mode to x. x = 1 for Weighing, x = 2 for Parts Counting, x = 3 forAnimal Weighing, x = 4 for Percent Weighing, x = 5 for Check Weighing, x = 6 for G/N/T,x = 7 for FillingxAW Set Animal Weigh Level to x. x = 1 for LOW, x = 2 for MEDIUM, x = 3 for HIGH.xAM Set animal mode. Where x is 1=Manual, 2=Semi and 3=AutoSAW Start Animal cycle.xT Download Tare value in current unit. Sending 0T clears tare.PID Print current user ID stringxID Program user ID string. 1-8 charactersAC Abort calibrationxUC Set user defined weightUC Initiate user calibrationIC Initiate internal calibrationPTIME Print current timemm/dd/yyDATE Set datehh:mm:ssTIME Set timePDATE Print current date5. TECHNICAL DATA5.1 RS232 CommandsField: Weight Unit Stab CR LFLength: Max 9 5 1 1 1