EN-32 ExplorerPro4.3 Error Codes ListError Codes ListThe following list describes the various error codes that can appear on the display and the suggested remedy.Data Errors1.0 Transient error (hardware error, probably static discharge). If error persists, the balance must beserviced.1.1 Balance temperature transducer hardware error.1.2 No data from main board.Tare Errors2.0 Balance is unable to stabilize within time limit after taring. Environment is too hostile or balance needsrecalibration.Calibration Errors3.0 Incorrect or no calibration mass used for calibration. Recalibrate with correct masses.RS232 Errors4.4 RS232 buffer is full.User Errors7.0 User entry out of bounds.7.2 Number outside of display capacity.Over-Under Load Errors8.0 Hardware error causing an internal weight signal which is too low. Check if pan is off. If not, the balancemust be serviced.8.1 Hardware error caused by an internal weight signal which is too high. Check load on the pan which maybe excessive. If error persists, the balance must be serviced.8.2 Power-on load out of specification (LFT only)8.3 Rated capacity exceeded. Remove excessive weight from pan.8.4 Underload condition on balance. Check that the proper pan is installed.8.5 Auto Cal weight internal sensor indicated its weight on the pan.CheckSum Errors9.1 Bad factory checksum. If error persists, have the balance serviced.9.2 Bad factory checksum. If error persists, have the balance serviced.9.3 Bad factory checksum. If error persists, have the balance serviced.9.4 Auto Cal data failed checksum. This failure will disable access to the Auto Cal feature (if installed).9.5 Factory calibration data failed checksum.9.6 Bad program checksum.9.7 Bad CMOS checksum.9.8 User calibration data failed checksum.9.9 Factory temperature compensation data failed checksum.