ExplorerPro EN-23TARGETWith the TARGET soft key highlighted, press the Enter button.The FILLING display is shown with the value of the Targethighlighted. Using the arrow buttons and the Enter button, theTarget value can be changed.FI LL ING 01/01/03 03:05:45 PMTA R GET S ET U P NEW TA R GET0.00 g0.00g*0%50% 90%110%100%Target: 0.00 gDiff: -0.00 g3.5 Balance SettingsThe Explorer Pro balance contains ten submenus that areaccessible from the Menu. The submenus are Calibration, BalanceOptions, Readout, Application Modes, Units, Interface, PrintOptions, GLP Print Options, Lockout and Factory Reset.Each of the submenus contain settings that will affect the operationof the balance. Please review all submenus settings to obtainthe best performance from the balance. Make the necessarysettings to suit specific needs.3.5.1 CalibrationPress the Menu button and select CALIBRATION. Press the enterbutton, the CALIBRATION screen is displayed. Explorer Probalances offers a choice of five calibration methods: InternalCalibration, Autocal Calibration, Span Calibration, LinearityCalibration and User Calibration,Linearity, Span and User calibration are disabled for TypeApproved/LFT balances Class II and Class III.Internal CalibrationOn Explorer Pro balances equipped with internal calibration,calibration is accomplished using the internal calibration mass.Internal calibration can be performed at any time providing thebalance has warmed up to operating temperature and is level.AutoCal: On, Off (default is On)Explorer Pro balances equipped with internal calibration alsocontain AutoCal. When AutoCal is set ON, the balance performsa self calibration when a measured predefined delta temperaturechange occurs. After calibration, the display returns to the lastapplication mode. Environmental conditions must be met for theAutoCal to be successfully completed.AutoCal Delta: -100 - +100 (default is 0)Explorer Pro balances equipped with internal calibration alsocontain AutoCal Delta. AutoCal Delta allows the internal calibrationmass value to be adjusted. This permits calibrating the balanceusing an external mass which is traceable to a certified standard.Perform the internal calibration procedure. After the calibration iscompleted, zero the balance.Place a certified mass equal to thespan calibration valuespan calibration valuespan calibration valuespan calibration valuespan calibration value of thebalance.IMPORTANT !DO NOT DISTURB THE BALANCEDURING CALIBRATION.CA LIB RATIONInternalAutoCal : OffAutoCal Delta (d): 0S panL inearityU serU ser Cal Weight (g): 500.00Cal TestExit