Keysight B2900 User’s Guide, Edition 5 4-51Front Panel ReferenceDisplay key groupDisplay Preference dialog boxThis dialog box is used to set the following parameters. The settings are saved to thenon-volatile memory.• V/I Limit for Each PolaritySelects the Limit (Compliance) setup operation.ON Individual setup ONPositive and negative limits can be set individually.OFF Individual setup OFF± Limit value is set.This function is available for the B2900 firmware revision 3.1 or later. And the“SWS” license is required to enable this function. See “To Use Functions otherthan Measurement” on page 5-5 to confirm the firmware revision and the licenseapplied to the B2900.Also see “File key group” on page 4-43 for the license.• Immediate V/I Update by KnobSelects the channel operation when the Source value or the Limit (Compliance)value is changed by using the digit pointer and the rotary knob.ON Real time update ONTurning the rotary knob changes the display value and updates thechannel output in real time.OFF Real time update OFFTurning the rotary knob just changes the display value. The value isapplied to the channel output by pressing the rotary knob.• Restore Trigger Mode on LocalStores the trigger mode setting on entering GP-IB remote state, ON or OFF.ON Saves the trigger mode setting in the non-volatile memory onentering GPIB remote state and restores the saved trigger modesetting when returning to GPIB local mode.OFF Disables this function.