Keysight B2900 User’s Guide, Edition 5 5-15Front Panel OperationsTo Set Source Output7. Press the rotary knob to change the pointer status to EDIT.8. Enter the pulse peak value (Peak), delay time (Delay), or pulse width (Width).9. Press the rotary knob or assist key to fix the setting value.10. Repeat steps 6 to 9 for all parameters.NOTE Applying the pulse voltage/currentPress the On/Off switch to start the Source value output. The Source value is used asthe pulse base value.Press the Trigger key to perform the specified pulse output and measurement.To Set Sweep OutputFor details on the sweep output function, see “Sweep parameters” on page 4-13 and“Sweep Output” on page 6-9.The following procedure sets the staircase sweep output.To set the pulsed sweep output, it is necessary to set the staircase sweep output andthe pulse output. For details on setting the pulse output, see “To Set Pulse Output”on page 5-14.1. Press the View key to display the Single view.2. Press the Show Sweep assist key to display the Sweep parameters.3. Move the field pointer to the Sweep Parameters field.4. Press the rotary knob to change the pointer status to EDIT.5. Press the LINEAR SINGLE, LINEAR DOUBLE, LOG SINGLE, or LOGDOUBLE assist key to select the relevant sweep operation. The pointer status ischanged to MOVE.6. Move the field pointer to the sweep parameter setup field (Start, Stop, Points, orStep).7. Press the rotary knob to change the pointer status to EDIT.8. Enter the sweep start value (Start), sweep stop value (Stop), number of sweepsteps (Points), or sweep step value (Step).9. Press the rotary knob or assist key to fix the setting value.10. Repeat steps 6 to 9 for all parameters.