3-4 Keysight B2900 User’s Guide, Edition 5InstallationInspecting the ShipmentInspecting the ShipmentPerform the following inspections when the Keysight B2900 and accessories arriveat your site.1. Before unpacking any component, inspect all boxes for any signs of damage thatmight have occurred during shipment, such as:• dents• scratches• cuts• water marksIf you suspect any damage, contact your nearest Keysight Sales and SupportOffice.2. When you open the boxes that contain the B2900 and accessories, check thecomponents against the contents lists attached to the boxes.If anything is missing, contact your nearest Keysight Sales and Support Office.3. Verify the operation of B2900 as described in “Checking the Operation ofKeysight B2900”.If any problem occurs, contact your nearest Keysight Sales and Support Office.Checking the Operation of Keysight B29001. Make sure that the standby switch is set to off.2. Connect the power cord from the B2900’s rear panel AC input connector(receptacle) to an AC power outlet at your site.3. Press the standby switch to turn on the instrument.The initialization screen will appear on the B2900’s front panel display, and thepower-on self-test is automatically executed.If the B2900 is operating normally, the front panel LCD displays as shown inFigure 3-1 or 3-2.