Labels 9Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 1231 Use a text editor to create each label. Each label can be up to ten characters inlength. Separate each label with a line feed.2 Name the file labellist.txt and save it on a USB mass storage device such as athumb drive.3 Load the list into the oscilloscope using the File Explorer (press [Utility] > FileExplorer).To reset the label library to the factory default1 Press [Utility] > Options > Preferences.2 Press the Defaul t Library softkey.This will delete all user-defined labels from the library and set the labels in thelibrary back to the factory default. However, this does not default the labelscurrently assigned to the channels (those labels that appear in the waveformarea).NOTE Label List ManagementWhen you press the Library softkey, you will see a list of the last 75 labels used. The list doesnot save duplicate labels. Labels can end in any number of trailing digits. As long as the basestring is the same as an existing label in the library, the new label will not be put in the library.For example, if label A0 is in the library and you make a new label called A12345, the newlabel is not added to the library.When you save a new user-defined label, the new label will replace the oldest label in the list.Oldest is defined as the longest time since the label was last assigned to a channel. Any timeyou assign any label to a channel, that label will move to the newest in the list. Thus, after youuse the label list for a while, your labels will predominate, making it easier to customize theinstrument display for your needs.When you reset the label library list (see next topic), all of your custom labels will be deleted,and the label list will be returned to its factory configuration.NOTE Pressing the Default Library softkey will remove all user-defined labels from the library andset the labels back to the factory default. Once deleted, these user-defined labels cannot berecovered.