Vertical Controls 3Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide 67When Fine adjustment is selected, you can change the channel's verticalsensitivity in smaller increments. The channel sensitivity remains fully calibratedwhen Fine is on.The vertical scale value is displayed in the status line at the top of the display.When Fine is turned off, turning the volts/division knob changes the channelsensitivity in a 1-2-5 step sequence.To invert a waveform1 Press the desired channel key.2 In the Channel Menu, press the Invert softkey to invert the selected channel.When Invert is selected, the voltage values of the displayed waveform are inverted.Invert affects how a channel is displayed. However, when using basic triggers, theoscilloscope attempts to maintain the same trigger point by changing triggersettings.Inverting a channel also changes the result of any math function selected in theWaveform Math Menu or any measurement.Setting Analog Channel Probe Options1 Press the probe's associated channel key.2 In the Channel Menu, press the Probe softkey to display the Channel ProbeMenu.This menu lets you select additional probe parameters such as attenuationfactor and units of measurement for the connected probe.The Probe Check softkey guides you through the process of compensatingpassive probes (such as the N2841A, N2842A, N2843A, N2862A/B, N2863A/B,N2889A, N2890A, 10073C, 10074C, or 1165A probes).