220 Keysight InfiniiVision 2000 X-Series Oscilloscopes User's Guide17 Waveform GeneratorThe Settings softkey opens the Waveform Generator Settings Menu which lets youmake other settings related to the waveform generator.See:• “To output the waveform generator sync pulse" on page 220• “To specify the expected output load" on page 220• “To use waveform generator logic presets" on page 221• “To restore waveform generator defaults" on page 227To output the waveform generator sync pulse1 If the Waveform Generator Menu is not currently displayed on theoscilloscope's softkeys, press the [Wave Gen] key.2 In the Waveform Generator Menu, press the Settings softkey.3 In the Waveform Generator Settings Menu, press the Trig Out softkey and turnthe Entry knob to select Waveform Generator Sync Pulse.To specify the expected output load1 If the Waveform Generator Menu is not currently displayed on theoscilloscope's softkeys, press the [Wave Gen] key.2 In the Waveform Generator Menu, press the Settings softkey.Waveform Type Sync Signal CharacteristicsAll waveforms exceptDC and NoiseThe Sync signal is a TTL positive pulse that occurs when the waveformrises above zero volts (or the DC offset value).DC N/ANoise N/A