Instruction ManualNivuFlow Mobile 600page 42 NFM 600 - rev. 02 / 10.07.201926 Connecting the Connector Box for Inputs/OutputsA Connector Box (Fig. 26-1 no. 6) must be used if you wish to additionally connect morethan one sensor or actuator using the inputs/outputs besides the flow velocity / level sensorson the NivuFlow Mobile 600 (Fig. 26-1 no. 1). For only one single external connection refer to chapter “25 Single externalConnection for Inputs/Outputs”.Important NoteThe Connector Box can either be fastened directly on the NivuFlow Mobile by using thehoop guards (Fig. 26-1 no. 2) or can be put/fastened to any other place.Fig. 26-1 NFM with Connector Box How to fasten the Connector Box on the NivuFlow Mobile:1. Remove the rubber buffers (not depicted) from the reverse side of the NivuFlowMobile (Fig. 26-1 no. 1). Both fastening plates (now loose, Fig. 26-1 no. 3) with-in the screw channels (Fig. 26-1 no. 4) are already required for the next step.2. Fasten the hoop guards (Fig. 26-1 no. 2) on the NivuFlow Mobile by using theaccompanying screws.3. Fasten the Connector Box (Fig. 26-1 no. 6) on the hoop guards by using theaccompanying screws.4. Connect the sensors to the Connector Box. See terminal wiring diagram (Fig.26-2).5. Plug (Fig. 26-1 no. 5) the connection cable from the Connector Box into themultifunction socket on the NivuFlow Mobile. For article numbers see chapter “52 Accessories”.