Parameter SettingNFM 600 - rev. 02 / 10.07.2019 page 7337 Parameters Menu ApplicationFig. 37-1 Application MenuThe following sections describe the general parameter setting procedures.Parameters for measurements using wet or clamp-on sensors are set in different ways. Theparameter setting procedures for the submenus >Measure Place< and >v-Paths< vary de-pending on the transit time mode.Due to this, both submenus will be described separately.37.1 Menu Measure PlaceThe >Measure Place< submenu is one of the most relevant basic menus when it comes tosetting parameters.The parameter set of a measurement place contains basic settings for:• Name of measurement place• Transit time mode• Arrangement and number of paths• Measurement medium and medium temperature• Type and dimensions of channel profile• Pipe material (incl. lining)• Possible solid sediments settings (sludge level)• Low-flow suppression• Measurement damping and stability37.1.1 Name of Measurement PlaceThis is where you can set the name of the measurement place.Default setting: “NIVUS1”.The default name is deleted automatically as soon as the first character of the new meas-urement place name is entered. Type the desired name of the measurement place into the text field and confirm with“Enter”. The name will be accepted and is shown in the main screen.