START-UPZEISS Mounting standard components Axioscope 5/7/Vario48 430035-7344-001 03/20183.1.7 Mounting the reflector insert• First, remove the cover cap (322HFig. 3-9/1) byinserting the Allen wrench (AF 3) into the slot(323HFig. 3-9/2). Turn the locking screwcounterclockwise and pull the cover capforward.• Push the reflector insert (324HFig. 3-9/3) with thereflector modules P&C (e.g. reflector turret with6 coded positions) into the upper part of thestand until it stops, hold it and tighten thelocking screw again.NOTEWhen the reflector insert is removed,the reflected light path is automaticallyblocked so that the glare will not affectthe operator.3.1.8 Mounting the mechanical stageDepending on the type, the mechanical stages caneither be stable or they can be rotated andcentered. The shift range is 75 mm in the X-direction and 50 mm in the Y-direction.Stages are available with a coaxial gear knob(325HFig. 3-10/7) located on the right or the left side ofthe stage.To assemble/disassemble a stage, remove the stagecarrier, then reattach the stage carrier with themounted stage to the stand (see section 3.1.15 onpage 59).(1) Stable mechanical stage, 75x50 Ra) How to remove the stage• Remove the four mounting screws (3 2 5 H Fig. 3-10/5)on the stage carrier (326HFig. 3-10/6) using the Allenwrench (AF 3) (327HFig. 3-10/4).• Remove the stage (328HFig. 3-10/1) from the stagecarrier. This is easier if you turn the stage and itscarrier over.Fig. 3-9 Mounting the reflector insertFig. 3-10 Changing the mechanical stage,75x50 R