START-UPAxioscope 5/7/Vario Mounting standard components ZEISS03/2018 430035-7344-001 573.1.12 Mounting the condenser carrier• Adjust the guide of the condenser carrier(391HFig. 3-21/3) using the coaxial knurled knob(392HFig. 3-21/2) until both screws (393HFig. 3-21/4)become accessible.• Mount the condenser carrier on the stagecarrier (394HFig. 3-21/1), slide it firmly and straightup to the upper stop and tighten both screws.3.1.13 Mounting the condenser• Slide the stage carrier with the focusing drive tothe upper stop.• Unscrew both centering screws (396HFig. 3-22/5) onthe condenser carrier until you cannot see theend parts anymore.• Unscrew the clamping screw (398HFig. 3-22/9) onthe right side of the condenser carrier until themaximum vertical adjusting range is usable.• Slide the condenser carrier (397HFig. 3-22/3) with theknurled knob for vertical adjustment(398HFig. 3-22/2) all the way down.• Open the front lens (if switchable) on thecondenser using the lever (395HFig. 3-22/7).• Insert the condenser (399HFig. 3-22/8) between thecondenser carrier (400HFig. 3-22/3) and the stagecarrier (401HFig. 3-22/1). Adjust the guide pin on thebottom of the condenser in the direction of theslot (402HFig. 3-22/6).• Push the condenser with the dovetail ring against the spring box (4 0 3 H Fig. 3-22/4) of the condenser carrieruntil the condenser can be placed on the condenser carrier horizontally.• Release the condenser and its pin screw will slide into the front slot (404HFig. 3-22/6).• Tighten the centering screws until they engage in the condenser’s dovetail ring.• Unscrew the clamping screw (398HFig. 3-22/9) without clamping the vertical drive.Fig. 3-21 Mounting the condenser carrierFig. 3-22 Mounting the condenser