EXPLORER BALANCES EN-103NOTE: The Information is dependent on the application and the ACCUMULATE setting.Examples are shown below.Mode Accumulate set OFF Accumulate set ONWeighing None N: 10Total: 10.000 kgAvg: 1.000 kgStd: 0.001 kgMin: 0.999 kgMax: 1.001 kgDiff: 0.002 kgCounting APW: 0.100 kg APW: 0.100 kgN: 10Total: 1000 PcsAvg: 100 PcsStd: 1 PcsMin: 99 PcsMax: 101 PcsDiff: 2 PcsPercent Ref Wt: 1.23 kg Ref Wt: 1.23 kgCheckWeighingUnder: 0.995 kgOver: 1.005 kgUnder: 0.995 kgOver: 1.005 kgN: 10Total: 10.000 kgAvg: 1.000 kgStd: 0.001 kgMin: 0.999 kgMax: 1.001 kgDiff: 0.002 kgDynamic Level: 0 Level: 0N: 10Total: 10.000 kgAvg: 1.000 kgStd: 0.001 kgMin: 0.999 kgMax: 1.001 kgDiff: 0.002 kg5.9.10 Print LayoutThis sub-menu is used to define the format of the data output to a printer or computer.FormatSet the printing format.MULTI = a multiple line printout is generated. A CRLF is added after each dataoutput.SINGLE = a single line printout is generated. A TAB delimiter is added betweeneach data output.FeedSet the paper feed.NONE = the paper remains in its current position after printing.LINE = move the paper up one line after printing.4 LINE = move the paper up four lines after printing.FORM = a form feed is appended to the output.5.9.11 Data Transfer FunctionOutput weighing results directly to a PC application. Setup is easy andno additional software is required.OFF = do not print.ON = print the specified settings.