EN-94 EXPLORER BALANCES5.5 Balance SetupEnter this sub-menu to customize Balance display functionality.5.5.1 Balance Set-up sub-menuStabilityIndicator RangeFilter Level Auto ZeroTrackingAuto TareGross Indicator Legal For Trade Graduations Date & TimeFactory default settings are shown below in bold.5.5.2 Stable RangeSet the amount the reading can vary while the stability symbol remains on.0.5 D = 0.5 graduations1 D = 1 graduation2 D = 2 graduations3 D = 3 graduations5 D = 5 graduationsNote: When LEGAL FOR TRADE (below) is set to ON, the setting is forced to 1 D.The setting is locked when the Security Switch is set to the ON position.5.5.3 Filter levelSet the amount of signal filtering.LOW = faster stabilization time with less stability.MEDIUM = normal stabilization time with normal stability.HIGH = slower stabilization time with more stability.5.5.4 Auto Zero TrackingSet the automatic zero tracking functionality.OFF = disabled.0.5 D = display maintains zero up to a drift of 0.5 graduationper second1 D = display maintains zero up to a drift of 1 graduation persecond.3 D = display maintains zero up to a drift of 3 graduations persecond.Note: When Legal for Trade is set to ON, the AZT setting is forced to 0.5 D.The OFF setting is still available. The setting is locked at the current setting when thesecurity switch is set to the ON position