EXPLORER BALANCES EN-143USB INPUTThe balance will respond to various commands sent via the interface adapter.Terminate the following commands when with a [CR] or [CRLF].Explorer Commands? print current mode0A turn Auto-print offSA Auto-print on, prints on stabilityCA Continuous Auto-print(n)A Auto-print on 1 to 3600 second intervals (n = 1 to 3600)C perform span calibrationL perform linearity calibration0M gram mode1M ounce mode2M troy ounce mode3M pennyweight mode4M parts counting mode5M pound modeT tare balance, same as pressing ON-ZEROV print software version(Esc)R reset balance to factory defaultsP same as pressing PRINTLE print last error code, i.e. [Err 0]0S print unstable data1S print stable data onlyP same as pressing PrintSP print stable weight onlyIP immediate print of displayed weight (stable or unstable)CP Continuous print of weightsSLP Auto-print stable non-zero weight onlySLZP Auto-print stable non-zero weight and zero readingxP Auto-print on 1 to 3600 second intervals (x = 1 to 3600)0P turns auto-print offPM print current modeM advance to the next enabled modePU print current unitU advance to the next enabled unitT same as pressing TareZ same as pressing ZeroPV print software versionAuto-Print OperationOnce Auto-Print is activated in the menu, the balance will send data as required.To temporarily stop Auto-Printing, press PRINT.If there is data in the print buffer the printer will finish printing this data.A second press will resume Auto-Printing.