EN-82 EXPLORER BALANCESContinue to enter the required information following the aboveprocedure:Positive Tolerance 1, via the +Tolerance 1 button.Negative Tolerance 1, via the –Tolerance 1 button.Positive Tolerance 2, via the +Tolerance 2 button.Negative Tolerance 2, via the –Tolerance 2 button.Tolerances can be in either weight (default) or percent –See Application Setup, Section 4.14.3.When finished, press Return to Application and the SQC homescreen appears.4.14.2 Begin the SQC ProcessTouch the BeginNew Batchbutton.First Batch: thescreen promptsfor the Batchname.Enter the Batchname, thenpress 9.The Process screen is then displayed.Follow the screen instructions, then press Acceptto store the sample weight.The current number of Samples, along with the Average, Sum,Minimum value, Maximum value, and Range are displayed.Continue to follow the screen instructions by weighing samplesand pressing Accept to store each sample’s weight.The sample size is 10 (default).It can be changed in Setup from 2 to 80 pieces.At any time, press End Batch to see the Individual resultsscreen, showing results for the current Batch.Once all the samples have been weighed,the Results screen appears automatically.If the COM interface is enabled, press Print to send the resultseither to a computer or a printer.Press Close to return to the current Batch process.A message offers the chance to save the batch, if desired.