EXPLORER BALANCES EN-47The Filling Application Setups (defaults in Bold)Item Available Settings CommentsEnabled On, Off Turns Application on/offSecondary Unit On, Off To show the Second Display lineStatistics Off, Manual, AutoLock Settings On, Off Press to lock the current setting(s) so theycannot be changed.Reference fieldsTarget Weight On, Off The Target weight is displayedDifference (Weight) On, Off Displays the Difference to the Target weightDifference (Percent) On, Off Displays the Difference to the Target weightGross On, Off Gross weight is displayedNet On, Off Net weight is displayedTare On, Off Tare weight is displayedPrint Options See Printing Section.Target Weight On, OffDifference (Weight) On, OffDifference (Percent) On, Off4.6.3 Statistics for FillingUse the Statistics function to compare a number of samples and examine the relative deviation of the samplesalong with other statistical data. A minimum of three samples is required.Statistics can be determined manually (Manual mode, with key press after each step) or automatically(Automatic Mode: weights are automatically recorded when stable). To use the statistics function, enable it firstin the Application Setup menu.Application SetupFirst, Enable StatisticsTouch the SetupIcon wrench (seebelow) to enterSetup menu.TouchStatistics.Choose eitherManual orAutomaticmode, thenpress Done.