EXPLORER BALANCES EN-1379.5 Communication9.5.1 Interface CommandsCommands listed in the following table will be acknowledged by the balance.The balance will return “ES” for invalid commands.TABLE 9-10. EXPLORER INTERFACE COMMAND LISTCommandCharacters 1) FunctionIP Immediate Print of displayed weight (stable or unstable). Attention: when LFT ON, IP couldonly print stable displayed weight.P1) Print displayed weight according to “Stable only” setting in the communication menu. Attention:when LFT ON, P could only print stable displayed weight.CP Continuous Print. Attention: when LFT ON, CP could not work.SP1) Print on Stability.SLP Auto Print stable non-zero displayed weight. Attention: the corresponding settings in thecommunication menu are modified too.SLZP Auto Print stable non-zero weight and stable zero reading. Attention: the correspondingsettings in the communication menu are modified too.xP Interval Print x = Print Interval (1-3600 sec) 0P ends interval Print. Attention: the correspondingsettings in the communication menu are modified too.0P See above.H Enter or get Print Header Lines. Attention: header text length is up to 25 characters, the settingformat is “H x “header line text””; and “H x” could get the corresponding header line in thebalance.Z Same as pressing Zero KeyT Same as pressing Tare Key.xT Establish a preset Tare value in displayed unit. X = preset tare value. Sending 0T clears tare (ifallowed).PT Prints Tare weight stored in memory.PM Print current application mode (weighing mode).xM Set current application mode to x. x depends on application, use application list 2) .M Scroll to the next enabled mode.PU Print Current weighing unit: g, Kg, lb, oz, etc….xU Set balance to unit x: g, Kg etc. x depends on unit list 3) .U Scroll to the next enabled unit.ON Brings out of StandbyOFF Goes to Standby.C Begin Span Calibration, same as trigger from calibration menu. Attention: when LFT ON, theoperation is not allowed.IC Begin internal Calibration, same as trigger from calibration menu.UC User Calibration (uses default weight), same as trigger from calibration menu. Attention: whenLFT ON, the operation is not allowed.AC Abort Calibration. Attention: when LFT ON, the operation is not allowed.xUC Set user defined weight and trigger one user calibration. Attention: the user defined weight isonly used for this command.