EN-84 EXPLORER BALANCES4.14.3 Application SetupThe Application can be customized for various user preferences.Touch the Setup Icon (wrench) to enter Application Setup fromthe home screen.The Setup Menu is now displayed.Select the list item to view or change the setting asdesired.To return to the Application home screen, touch Done.The SQC Application Setups (defaults in Bold)Item Available Settings CommentsEnabled On, Off Turns Application on/offNumber of Samples 2 – 80 (default 10) To select the number of samples in a BatchTolerance Mode Weight,PercentTolerance used during measurementsAuto Sample Mode Manual,Semi Automatic,AutomaticTo select the mode of operationLock Settings On, Off Press to lock the current setting(s) so theycannot be changed.Reference fieldsNumber of Batches On, Off The total batchesNumber of Samples On, Off The total samplesAverage On, Off The Average value is displayedTotal On, Off The Sum value is displayedMaximum On, Off The Maximum value is displayedMinimum On, Off The Minimum value is displayedRange On, Off The Range is displayedPrint Options See Printing Section.Number of Batches On, OffNumber of Samples On, OffAverage On, OffTotal On, OffMaximum On, OffMinimum On, OffRange On, Off