PUTTING THE INSTRUMENT INTO OPERATIONAxio Vert.A1 Mounting Standard Components Carl Zeiss05/2012 431030-7044-001 513.1.10.2 Changing the Halogen Lamp forTransmitted LightWARNINGWhen changing the halogen lamp,always be sure to allow for sufficientcooling of the lamp and to pull thepower supply plug.CAUTIONDo not touch the bulb with yourbare hands. If required, clean thebulb with pure alcohol before youswitch it on for the first time inorder to prevent dirt from burninginto the glass.• Pull the illuminator cap (Fig. 3-14/1) upward offthe carrier support (Fig. 3-14/5).• Pull the malfunctioning halogen lamp(Fig. 3-13/2) out of the mount (Fig. 3-13/3).• Insert the new halogen lamp (Fig. 3-13/2) intothe mount (Fig. 3-14/3). In so doing, make surethat the tip of the mount engages in thecentering notch of the carrier plate (see arrowin Fig. 3-13) of the halogen lamp.• Plug the illuminator cap (Fig. 3-14/1) onto thecarrier support (Fig. 3-14/5). For properalignment, slide the notch in the cap onto thealignment stud (Fig. 3-14/4).• Reconnect the power plug.Fig. 3-14 Changing the Halogen Lamp forTransmitted Light