PUTTING THE INSTRUMENT INTO OPERATIONAxio Vert.A1 Mounting Standard Components Carl Zeiss05/2012 431030-7044-001 533.1.11.1 Attaching the Fluorescence Illuminator for Reflected LightFor safety reasons, the HBO 50/HBO 100 fluorescence lamps and the HXP 120C mercury vapor short-arclamps are packaged and delivered to the customer separately.Before those lamps can be put into operation, they must, therefore, be inserted into the illuminatorhousings.The procedure to follow when installing or replacing the lamp is described in the operating instructionsenclosed with the devices.To install the illuminators, proceed as follows:• Loosen the clamping screw (Fig. 3-16/3) and remove the protective cap from the reflected light port(Fig. 3-16/2).• Insert the HBO 50 (Fig. 3-16/1) or the adapter for the HXP 120C illuminator with the dovetail directlyinto the reflected light port (Fig. 3-16/2) and lock it into position with the clamping screw (Fig. 3-16/3),using ball-headed screwdriver SW 3.• When it comes to the HBO 100, be sure to attach the 35-mm distance ring (Fig. 3-16/4) onto thereflected light port in the same manner before mounting the illuminator (Fig. 3-16/5).• Connect the multipole illuminator plug (Fig. 3-16/6 or 7) to the socket of the appropriate powersupply unit and secure it with the retaining ring.• First connect the power supply cable to the HBO 50/HBO 100 power supply unit and then to thepower socket.• Use the ON/Off switch (Fig. 3-16/9 to turn on the power supply unit for the HBO 50 or the ON/Off(Fig. 3-16/8) to activate the HBO 100.Fig. 3-16 Attaching the HBO 50 Illuminator or the HBO 100 Illuminator