Adjustment and Calibration IOLMaster 700Instructions: (Step 2)(1) Wait, till output-window indicates thatwork is done.(2) Check the ‘Console-Output’ for warningsand/or error messages.(3) Click 'Finish' button.Fig. 16 – S301 Exchange complete measurement head(3)Attention!The factory-adjusted device-settings must be verified against the local test eye.For this reason, it’s recommended to close the Service Tool, restart the device andcomplete a calibration check using the IOLMaster 700 application!S401 – Exchange complete computer base with OS & application pre-installedDescription:Use this function after the computer basehas been replaced and both disk-image anduser-data have been restored.This use-case completes the following tasks:(1) Set the serial-number of the newcomputer base.(2) Retrieve device-settings from bothmeasurement head and local files.(3) Join both sets of device-settings:Use serial-number of computer base frommanual input,use all other device-settings frommeasurement-head (mirror).(4) Persist the joined settings to bothmeasurement-head and local-files.(5) Backup the joined settings onto theservice-drive (mSata).Fig. 17 – Exchange complete computer base with OS &application pre-installed (1)Service Manual Confidential and Proprietary SM-30-4060Order No. 2075-325 4 - 10 Version 03, 2014-10-13