IOLMaster 700 Adjustment and CalibrationCleaning the Test EyeMove the Test eye out of the PI with the1)service tool. Refer to 4.3.2, page 5.Clean the test eye.2)Return the Test eye into the PI with the service3)tool. Refer to 4.3.2, page 5.Fig. 28 - Cleaning the test eyeCleaning the Keratometry AssemblyFig. 29 - Front glass Fig. 30 - Keratometry lenses (18)Cleaning the Optical surfaces inside the Instrument headPreconditions: Instrument Head covers removed. Refer to section 3.2.1. Polystyrene dust covers removed. Refer to section 3.2.2. Service tool is running. Refer to section 4.3.2, page 5- Use case 705NoteIt is recommended to perform inside the head cleaning with compressed-air, as theinside components normally are only contaminated with dust (no grease contaminants).If using canned compressed air, do not allow liquid propellant to strike optical surfaces,as spotting may result on the surfaces. If liquid cleaning is necessary (e.g. fingerprints),proceed as showed in the following sections.SM-30-4060 Confidential and Proprietary Service ManualVersion 03, 2014-10-13 4 - 15 Order No. 2075-325