Adjustment and Calibration IOLMaster 7004.4.2 Paddle PolarisationThe Paddle Polarisation procedure allows accurate adjustment of the paddle position to ensuremaximum SNR during measurement. The paddle polarisation tool will be automatically removed afterrestarting the instrument, therefore it must be installed each time the procedure is performed.Procedure:Enter the service mode1)See section 4.2.1, page 3.Ensure the File Based Write Filter protection2)is ENABLED. See section 4.5.2, page 29.Fig. 44 - File Based Write Filter EnabledUsing Windows Explorer, navigate to the3)folder: Y:\PolarizationToolFiles.Run (double click) the4)uEye_LabView_40200.exe.Fig. 45 - Run the uEye_LabView_40200.exeSelect “Englisch” and press “weiter >”.5)Press “Next >”, press “Next >” and press6)“Next >”.Press “Install”.7)Fig. 46 – Install the file12Service Manual Confidential and Proprietary SM-30-4060Order No. 2075-325 4 - 20 Version 03, 2014-10-13