Troubleshooting Error Codes IOLMaster 7002030Invalid operation while getting a cameraparameter2031A general timeout occurred. A processcouldn't be terminated or succededwithin the specfied period2032Device Not Available: The initializeddevice was disconnected2033A camera was removed. This isindependent of the device handle2034An error occured while image capturing isrunning (common/undefined)2035Not enough destination memoryallocated or all destination buffers lockedby the application:- Release locked destination memory- Allocate more destination memory- Reduce the frame rate so that there ismore time to process the filled destinationmemory2036The current image could not be processedcorrectly. It's an internal error duringinternal processing of the image2037The destination buffers are locked andcould not be written to2038No free internal image memory isavailable to the driver. The image wasdiscarded2039The camera is no longer available. It is notpossible to access images that havealready been transferred2040USB Transfer Error: The image was nottransferred over the USB bus2041The maximum allowed time for imagecapturing in the camera was exceeded2042 Wrong IDS API driver version2043 There is no device available or conntected2044There is more than one device available orconntected2045The process is running under 32-bit(instead of 64-bit) or wrong windowsversion (must be 64-bit)Service Manual Confidential and Proprietary SM-30-4060Order No. 2075-325 5 - 6 Version 03, 2014-10-13