Troubleshooting Error Codes IOLMaster 7001009An error occured during uploading ascanpattern-file intothe device-head.1010A specific scanpattern expected to beloaded onto the device-headis either not present there or is out ofdate.1011A CsaPixyException has been catched andrethrown as HalException.The origign exception is provided asInnerException of the HalException.This error is a PI-fault.- Change the PI or repair the connection1012Not used until yet (similar toCsaPixyError).1013Indicates that a chinrest-moving-command cannot be executed because arequired reference-drive of the chinresthas not completed yet.1014The requested command cannot beexcuted because HAL is in faultedstate due to a preceeding HAL-error.1015The requested command cannot beexcuted because the HAL has not beeninitialized until yet.1016The caller tried to initialize the HAL asecond time although initializingalready has been done or just is inprogress.1017A HAL-operation has succeeded, butwithin the callers callback-methodoccured an exception.1018The requested HAL-command is invalid,invalid at this time or currentlynot support. For example calling'DoSingleOctScan' while in operating-mode'Idle' causes this error.1019Not used until yet. Would indicate aoverflow of the HAL-command queue.1020A HAL-command has neither completednor failed within a command-specifictimeout-span. The timeout can be aexecution- or a transmission-timeout.Service Manual Confidential and Proprietary SM-30-4060Order No. 2075-325 5 - 4 Version 03, 2014-10-13