Adjustment and Calibration IOLMaster 700BACKUP (mSATA SSD 32 GB)Bootable drive containing the ‘Acronis True Image Personal’ application for disk backup/recovery.Paddle Polarisation tool files Y:\PolasizationToolFiles\ (See section 4.4.2, page 20 for further details).Config files (backed up after using the Service Tool): Y:\DeviceConfiguration_Backup\ (config, device.settings.xml, filterGenerator.settings.xml)Acronis Images: Y:\Factory Image\*.tib (factory image of System and Data drives)STM32_PUB (EEPROM 68KB)Windows can not be used for copy/cut/paste of files into the measurement head (STM_PUB32).Measurement head configuration files can be found on both the computer base (System) and themeasurement head (STM_PUB32) – they must be identical (this can be confirmed with the servicetool)! Z:\config (binary configuration file = C:\DeviceConfiguration\config) non-visible/accessible: Z:\device.settings.xml non-visible/accessible: Z:\filerGeneartor.settings.xmlService Manual Confidential and Proprietary SM-30-4060Order No. 2075-325 4 - 28 Version 03, 2014-10-13