IOLMaster 700 Adjustment and Calibration4.5 Adjustments / Alignments of the Computer Base4.5.1 HDD Structure (Databases, log files, OS etc.)Hardware Disk Label Location Connection BIOS PortSSD 32GB System (C:\) Computer Base SATA P1HDD 500GB Data (D:\) Computer Base SATA P5mSATA SSD 32GB BACKUP (Y:\) Computer Base mSATA P2EEPROM 68KB STM32_PUB (Z:\) Meas. Head USB IBIS_STM32E_RamDiskSystem (SSD 32GB)Protected by File-Based Write Filter (FBWF) preventing any user/malicious software from makingchanges to the file system. The FBWF can be disabled/enabled following the steps in 4.5.2, page 29.Important files/folders: Operating System (OS) – Windows Embedded Standard 7 SP1 C:\Program Files\CZM\IOLMaster700\ (Application software) C:\Program Files\CZM\IOLMaster700ServiceTool\ (service tool) C:\Program Files\CZM\IOLMaster700ServiceTool\scripts (scripts for deleting the databases) Microsoft SQL server (used for application database management)Not Important: C:\Program Files (x86)\Gamut\ (log viewer software) C:\Windows\System32\fbwfmgmt.exe (File based write filter manager)Config files: C:\DeviceConfiguration\config (binary configuration file) C:\DeviceConfiguration\device.settings.xml (all meas. head configuration values) C:\DeviceConfiguration\filterGenerator.settings.xml (all meas. head filter configuration values)Data (HDD 500GB)Database files: D:\Data\DB\CzmDefault.mdf (Primary database; includes patient data, measurement results, lensdata) D:\Data\DB\ CzmUserManagement.mdf (User database; includes user data and role) D:\db.default\storageroot\ (all patient data image files) D:\db.usermanagement\ (nothing useful is stored here)Logs: D:\Logs\ (includes application log files, NIM log files etc.)Not Important: D:\PDF_Export\ (shared folder where the PDF printouts are stored) D:\Settings\ (xml files with application settings and each user specific settings ie. What isdisplayed in the Analysis, is Reference Image taken)SM-30-4060 Confidential and Proprietary Service ManualVersion 03, 2014-10-13 4 - 27 Order No. 2075-325