User manual SEFELEC 5x Series 154 v1.06DEFINITION OF INPUT-OUTPUT SIGNALSCOMIN (C6-5)Electrical common line for the input signals.COMOUT (C7-4)Electrical common line for the output signals.CTRLIN (input C6-6)Request to take over control of the measuring instrument by the PLC.MES_DCH (input C7-2)Switching to measurement or discharge.In the high logic state: Measurement.In the low logic state: Discharge.TYPE (input C7-1)This input corresponds to the previous choice of the measurement function of the XSrange, with the SEFELEC 5x range:At the low logic state: offsets the memory numbers of 256 positions (256-512)At the high logic state: does not offset the memory numbers. (0-256)N0, N1, N2, N3 (inputs C6: 7,8,9,10)N4, N5, N6, N7 (inputs C8: 1,2,3,4)Binary encoding enabling the user to select a memory number (0 to 255) and an associat-ed settings sequence file. See also TYPE inputN7 N6 N5 N4 N3 N2 N1 N0 Memory File name0 0 0 0 0 0 0 0 0 Name10 0 0 0 0 0 0 1 1 Name20 0 0 0 0 0 1 0 2 Name30 0 0 0 0 0 1 1 3 Name40 0 0 0 0 1 0 0 4 Name50 0 0 0 0 1 0 1 5 …0 0 0 0 0 1 1 0 6 ….0 0 0 0 0 1 1 1 7 ….0 0 0 0 0 1 0 0 8 ….0 0 0 0 1 0 0 1 9 ….…. …. …. …. …. …. …. …. …. ….1 1 1 1 1 1 1 1 255 Name256The file names only correspond to Sequence files. To run a measurement function alone, put acall line only for the desired function in the Sequence.