User manual SEFELEC 5x Series 182 v1.06OHM (OHMmeter) :Sets the main thresholds unit (Ohm) and the ground continuity display function.Caution : The action on this command will erase all previous recordedthresholds (Reset to zero).VOLT (VOLTmeter) :Sets the main thresholds unit (Volt) and the ground continuity display function.Caution : The action on this command will erase all previous recordedthresholds (Reset to zero).HTIM (HTIMe) :Sets a new value in second to the current hold time parameter to the parameter setof the running function.For instance : HTIM 3.2RTIM (RTIMe) :Sets a new value in second to the current rise time parameter to the parameter setof the running functionFor instance : RTIM 10.5FTIM (FTIMe) :Sets a new value to the current fall time parameter to the parameter set of the run-ning function.For instance : FTIM 0.5HLIM (HLIMit) :Sets a new value to the high limit parameter to the parameter set of the runningfunction.For instance :MEG:HLIM 2.0E+6 (new value in Ohms)HIP:HLIM 1.45E-4 (new value in Amps)GND:HLIM 1.00E-1 (new value in milliohm)The unity is automatic, given in Volt, Ohm or Amp and function dependent.CB HLIM :Same command as HLIM, but reset the low limit parameter before to change thehigh limit value.LLIM (LLIMit) :Sets a new value to the low limit parameter to the parameter set of the runningfunction.For instance :MEG:LLIM 1.0E+6 (new value in Ohms)HIP:LLIM 3.50E-6 (new value in Amps)GND:LLIM 5.02E-2 (new value in milliohm)The unity is automatic, given in Volt, Ohm or Amp and function dependent.TIM (TIMe) AUT/PERMSelect the temporization mode. AUT for automatic, PERM for Permanent test untila fault occurs.