User manual SEFELEC 5x Series 34 v1.06INSTALLATION OF THE DIFFERENT CONNECTORS AND CORDSBefore powering the measurement device, ensure that the following accessories are installed,and their plugs locked:• The C5 safety plug (10 points) must be installed on the socket marked C5 (caution, there ismechanical fool-proofing on the sockets)• Connect the mains power cord to the rear panel and to a 16A wall socket with groundIMPORTANTIf the mains power cord is replaced, be sure to always use a mains cordsuitable for the power and CE certified.INSTRUCTIONS FOR FITTING IN THE BAYIn cases when the SEFELEC 5x range device needs to be fitted in a bay, follow the instructionsbelow:The dimensions of the device are: height 3U, width to the 19'’ standard, depth 520 mm with theconnectors. Use bays to the 19'' standard with a depth of 600 mm minimum.1 - Take the references of the bay and use the in-bay mounting kit supplied by the brand.For a device with a height = 3U the kit is usually composed of 2 runners and 4 screws(Caution: take the runners to suit the depth of the bay).2 - Preparation of the device: Remove the 4 feet by unscrewing the 4 screws (hexagon sockethead cap screws). Then remove the 2 screws on each side that hold the cover on the front of thedevice (Torx socket head cap screws).See below:3- Equip the device of the SEFELEC 5x range with the kit: SEFA-KR. It consists of 2 bracketsand 4 screws (countersunk torx head).