User manual SEFELEC 5x Series 19 v1.06MEASUREMENT SPECIFICATIONSDielectric strength, 50 VA(SEFELEC 56-H/D/S)Output voltageAdjustable from 0.100 to 5.000 kV in ACSinusoidal alternating with frequency of 50 Hz or 60 HzAdjustable from 0.100 to 6.000 kV in DC(+) pole in DC to groundResidual DC ripple < 1% for a current < 3 mAStability < ±3% for mains ΔV of ±10%Accuracy of the generation in relation to the setpoint for voltages between 100 and 5,000 V AC(6,000 V DC) and a current flow less than 3 mA± (2%+5 V)Maximum capacity of the sample measured < 1 μF (discharge time < 10 sec.)Discharge of the element tested and internal ca-pacitors1.5 MΩ in DCVoltage readingBy a voltmeter connected directly to the output terminals.Accuracy ± (1.5% +5 V)Resolution 6,000 pointsNominal currentfrom 800VAC to 5000VAC 10 mAfrom 400VDC to 6000VDC 8 mAShort-circuit currentFor the adjustment of the maximum voltage in AC < 20 mAFor the adjustment of the maximum voltage in DC < 20 mAFault detection:In current variation ΔIThe ΔI (delta I) sensor automatically performs the subtraction between the current flowing nor-mally in the sample under test (I = U/Z) and the one that occurs suddenly when a fault occurs:(I' = I + I fault).Amplitude adjustable from 100 μA to 10 mA ± 10% in steps of 100 μAPulse width of 10 μS ±20%.