User manual SEFELEC 5x Series 22 v1.06Dielectric strength, 500 VA(SEFELEC 506-H/D/S)Output voltageAdjustable from 0.100 to 5.000 kV in ACSinusoidal alternating with frequency of 50 Hz or 60 HzAdjustable from 0.100 to 6.000 kV in DC(+) pole in DC to groundResidual DC ripple < 1% for a current < 3 mAStability < ±3% for mains ΔV of ±10%Accuracy of generation in relation to the setpoint for voltages between 100 and 5,000 VAC (6,000 V DC) and a current flow less than10 mA± (3%+5 V)Maximum capacity of the sample measured < 1 μF (discharge time < 10 sec.)Discharge of the element tested and internalcapacitors1.5 MΩ in DCVoltage readingBy a voltmeter connected directly to the output terminals.Accuracy ± (1.5% +5 V)Resolution 6,000 pointsNominal currentfrom 800VAC to 5000VAC on capacitive load >100 mAfrom 1500VAC to 5000VAC on resistive load >100 mAfrom 400VDC to 6000VDC >20 mAShort-circuit currentFor adjustment of the maximum voltage in AC >= 200 mAFor adjustment of the maximum voltage in DC >= 100 mAFault detection:In current variation ΔIThe ΔI (delta I) sensor automatically performs the subtraction between the current flowing nor-mally in the sample under test (I = U/Z) and the one that occurs suddenly when a fault occurs:(I' = I + I fault).Amplitude adjustable from 1 mA to 100 mA ± 10% in steps of 1 mAPulse width of 10 μS ±20%.