User manual SEFELEC 5x Series 16 v1.06The dielectric strength tests are performed at voltages adjustable from 100 VAC to 5,000 VACand 100 VDC to 6,000 VDC. The rated current is 10 mA for the 50 VA models and 100 mA for500 VA models. Breakdown voltages and currents are stored after circuit break and cut-off of thehigh voltage.The insulation resistance is measured at voltages of -20 VDC to -1,000 VDC.The device gives, in direct readings with display of the units, resistance values of100 kΩ to 200 GΩ (2 TΩ on option).The ground continuity resistance is measured at currents of 5 to 32 A AC (50 A AC on option)with open circuit voltages of 6 or 8 V. The device gives, in direct readings with display of theunits, resistance values of 1 mΩ to 1,000 mΩ,The devices in the SEFELEC 5x range are equipped as standard to communicate with any sys-tem provided with an API, RS232, Ethernet or even IEEE 488 (optional) interface.OPTIONS (TO BE ORDERED ACCORDING TO THE REQUIREMENT)Reference DescriptionSEFO-IEEE488 IEEE 488-2 interface optionSEFO-CO160 red/green lamp for presence of high voltage optionSEFO-5X50A 50 A AC for ground continuity optionSEFO-5XREAR rear outputs optionSEFO-5X3MY short-circuit current limit of < 3 mA AC and DC in HV 50VA optionSEFO-5x2TB 2 TOhm in insulation optionSEFO-5XRC remote control for accessories option