User manual SEFELEC 5x Series 181 v1.0618.1.4. COMMANDS DEPENDENT ON THE DEVICEMEG (MEGohmmeter) :Selects the megohmmeter. Must be sent from the initialization display.HIP (HIPot) :Selects the Hipot function. Must be sent from the initialization display.GND (GrouND) :Selects ground continuity function. Must be sent from the initialization display.SEQ (SEQuence) :Selects the sequence function. Must be sent from the initialization display.CONF (CONFig) :Selects the configuration function of the unit. Must be sent from the initializationdisplay.PAR (PARameter) :Selects the parameter set for the running function.i.e.: MEG : PAR 1DCV (DCVoltage) :Sets a new value in volt to the current DC voltage parameter to the parameter setof the running function.i.e. : DCV 500ACV (ACVoltage) :Sets a new value in volt to the current AC voltage parameter to the parameter setof the running function without changing the pre-set frequency.Hipot tester : Min. : 100 Max. : 5000For instance : hipot : ACV 5000ACV50 (ACVoltage) :Sets a new value in volt to the current AC (50Hz frequency) voltage parameter tothe parameter set of the running functionHipot tester : Min. : 100 Max. : 5000For instance : hipot : ACV50 5000ACV60 (ACVoltage) :Sets a new value in volt to the current AC (60Hz frequency) voltage parameter tothe parameter set of the running functionHipot tester: Min.: 100 Max. : 5000For instance: hipot : ACV60 5000ACC (ACCurrent) :Sets a new value in ampere to the AC current parameter to the parameter set ofthe ground continuity function.