User manual SEFELEC 5x Series 24 v1.06Dielectric strength, 500 VAMeasurement of the ACTUAL permanent currentThe reading of the real current, linked to the resistance, is obtained through a shunt located di-rectly in the test circuitResolution 11,000 pointsAccuracy0.01 mA to 110.00 mA AC ±(3% + 1 mA)Signalling of a faultBy a PASS or FAIL colour message on the LCD display and an audible signalWith memory of Breakdown voltage.Current in case of a fault (IMAX mode)In the event of a fault, the high voltage cut occurs without overvoltage.Time delayMANUALNo time is applied to the test or the measurement. The control of the voltage is manual by actionon the up and down arrows of the LCD screen. The test or the measurement is stopped only if afault occurs or if the user requests to stop the test by pressing the red button on the front panel.PERMANENTA rise time applies to the test and then the output voltage is equal to the setpoint. The test or themeasurement is stopped only if a fault occurs or if the user requests to stop the test by pressingthe red button on the front panel.AUTOIn AUTO mode: The test consists of 3 successive phases during which the voltage increaseslinearly up to the desired value (Rise), is maintained at the programmed value (Hold) and is thengradually reduced to 0 (Fall)Rise and Fall times programmable from 0.0 to 9999.0 sec. in steps of > 0.1 s.Hold programmable from 0.1 to 9999.0 sec. in steps of > 0.1 s.Accuracy +/- 20 msec.