CWS-BA-e-0610 13Mass terminal7.2.1 PlacingLevel balance with foot screws until the air bubble of the water balance is in the pre-scribed circle.7.2.2 Scope of deliverySerial accessories:• Balance• Weighing plate• Detachable display part• Mains power supply• Operating Manual7.3 Mains connectionPower is supplied via the external mains adapter. The stated voltage value must bethe same as the local voltage.Only use original KERN mains adapters. Using other makes requires consent byKERN.Connect the housing to earth, if there is a suspicion that static electricity could pose aproblem.7.4 Initial CommissioningIn order to obtain exact results with the electronic balances, your balance must havereached the operating temperature (see warming up time chap. 1). During this warm-ing up time the balance must be connected to the power supply (mains, accumulatoror battery).The accuracy of the balance depends on the local acceleration of gravity.Strictly observe hints in chapter Adjustment.Mains adapter