CWS-BA-e-0610189 Parts countingWith parts counting you can either count parts into a container or remove parts froma container. To count a greater number of parts the average weight per part has tobe determined with a small quantity (reference quantity). The larger the referencequantity, the higher the counting exactness. High reference must be selected forsmall parts or parts with considerably different sizes.9.1 Start of add-up processTurn on balance by pressing the ON/OFFkey. If the display is not on zero, press the-key.The three LEDs on the keys"piece weight by" are flashing. This is arequest to enter the method you wish toapply for entering the reference weight.a00.000 Kg 0.00000 gThrough the selection of the keys the following options are available to enter or retrievethe reference weight:Retrieving a stored referenceweight via ID no. (chpt.. 9.4)Numeric input of referenceweight (Chpt.. 9.3)Determination of referenceweight by weighing (chpt.. 9.2 )Information:You can return to this point during operation at any time by pressing the CE keywithout deleting your settings (tare value, M+, tolerance limits).