Language Dictionary - 4[SOURce:]CURRentChannel SpecificThis command sets the current that the load will regulate when operating in constant current mode. Referto Table 4-1 for model-specific programming ranges.Command Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] Parameters 0 through MAX | MINimum | MAXimumUnit A (amperes)*RST Value MINimumExamples CURR 5 CURR:LEV .5Query Syntax [SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]?Returned Parameters Related Commands CURR:TLEV CURR:MODE[SOURce:]CURRent:MODEChannel SpecificThis command determines whether the current settings are controlled by values in a list or by theCURRent command setting.FIXed The current settings are determined by the CURRent command.LIST The current settings are determined by the active list.Command Syntax [SOURce:]CURRent:MODE Parameters FIXed | LIST*RST Value FIXedExamples CURR:MODE FIXQuery Syntax [SOURce:]CURRent:MODE?Returned Parameters Related Commands CURR: CURR:TRIG[SOURce:]CURRent:PROTectionChannel SpecificThis command sets the soft current protection level. If the input current exceeds the soft currentprotection level for the time specified by CURR:PROT:DEL, the input is turned off. Refer to Table 4-1 formodel-specific programming ranges. (Use CURR:PROT:DEL to prevent momentary current limitconditions caused by programmed changes from tripping the overcurrent protection.)NOTE: A fixed offset equivalent to 1% of the unit’s full-scale rated current is automatically added to theprogrammed value. So, for example, if you are programming the current protection of anN3306A to 100A, the actual setting will be 101.2A (programmed value + 1% of rated current).Command Syntax [SOURce:]CURRent:PROTection[:LEVel] Parameters 0 through MAX | MINimum | MAXimumUnit A (amperes)*RST Value MAXimumExamples CURR:PROT 10Query Syntax [SOURce:]CURRent:PROTection?Returned Parameters Related Commands CURR:PROT:DEL CURR:PROT:STAT61